Clinical chemical analyser
 |
HA-1900
|
|
Features
- Biochromatic 7 wavelength optical
system
- Pre-programmed calculations for
kinetic, two point and end-point assays
- Large color LCD display, Windows CE
compatible operating system
- Lamp save feature
- Temperature controlled cuvette well :
25℃, 30℃, 37℃
- Large memory to store 160 test
routines (47 preprogrammed) and up to 600 sample results
- User definable units
- Reaction curve display
- Quality control feature
- Multi-format result output including
comprehensive patient report
- Built-in modem to support
long-distance maintenance and software upgrades(optional)
Please email or telephone for prices and availability
|
|
|
|
|
System Specifications
|
Principle |
|
Absorbance |
Wavelengths |
|
340, 405, 500, 546, 578, 620, 670nm, plus one more
optional filter available |
Display-range |
|
0.000-3.000 Abs |
Measurement-range |
|
0.000-2.500 Abs |
Resolution |
|
0.001 Abs (display), 0.0001 Abs (calculation) |
Half band width |
|
8nm + or - 1nm |
Linearity |
|
r>=
0.999 |
Flow cell |
|
30μl
quartz flow cell |
Carry over |
|
< 1% |
Minimum volume |
|
200μl |
Maximum volume |
|
3000μl |
Temperature controller |
|
Peltier 25℃, 30℃,
37℃ + or - 0.1℃ and ambient temperature |
Stability |
|
No more than 0.003A/hour |
Display |
|
7" colour LCD (640 x240 resolution, 256 colors) |
Input |
|
External mouse and keyboard |
Data Storage Capacity |
|
160 test routines (47 preprogrammed), up to 600
sample results |
Output |
|
Built-in printer or external parallel printer |
Data Communication |
|
RS-232 serial cable |
Dimensions L × W × H |
|
360 x 300 x 170mm |
Weight |
|
7.5 kg |
Power supply |
|